Main navigation

Sciemetric

The Sentinel I28 single-station leak detection system is designed for easy operator setup and troubleshooting. It features a small footprint (12” x 9.25” x 8.75”) for easy installation in limited space, including wall mounting. All CTS instruments include password protected program and test result storage.

  • Powerful 32-bit processor with 24-bit A/D converter
  • Language-neutral color LCD with 12 line, 480 x 272 operator display
  • XY graphing (pressure, loss, or flow)
  • English, Chinese, Korean, Spanish language options
  • Indicator light bar identifies test status: Pass/Fail/In-test
  • Expanded instrument control through EtherNet/IP™ communication
  • Barcode-ready input RS232 alpha-numeric string stored with test results
  • Modular pneumatic and control packages
  • Auto calibration includes mass flow leak standard certified for pressure and flow
  • Transducer zero and span are automatically reset during auto calibration
  • “Performance factor” calibration feedback ensures accurate results
  • Manually configurable calibration program
  • Self-test isolates the Sentinel I28’s pneumatics to verify a “no leak” condition
  • Firmware validation document included

Standard pressure transducer, valve manifolds, and regulator ranges:

 -14.5, -2, -0.1 to 4” HG, 0.5, 2, 30, 100, 200, 400, 500, or electronic regulation (PSIG)

  • Optional low volume test manifold available
  • Optional high cV flow valves and manifold available up to 100 psig range allows 5X faster fill and exhaust to reduce cycle times on large parts
  • Optional low volume (0.07cc) manifold available
  • Absolute pressure transducer
  • Six point pressure calibration
  • Data resolution 0.00001 psi/ and scc/m.
  • Optional electronic regulator available with multi-point calibration

CTS will optimize the flow sensor and flow range of your Sentinel I28 leak testing equipment to meet the needs of your unique application. Contact our application engineering team for available mass flow sensors.

Units of measure can be instrument global or per test program

Pressure

  • PSIG
  • PSIV
  • inH2O
  • cmH2O
  • mmH2O
  • kg/cm2
  • ATM
  • inHg
  • cmHg
  • mmHg
  • kPa
  • Pa
  • MPa
  • Bar
  • mBar
  • Torr

Flow

  • sccm
  • sccs
  • scch
  • slpm
  • slps
  • slph
  • s

Time

  • sec
  • msec
  • min

 

 

 

 

  • Expanded control through EtherNet/IP™ communication
  • 12 inputs and 12 outputs, with up to five (5) programmable tooling motions with feedback
  • Part marking control
  • Tooling I/O is externally connected
  • I/O 24 VDC: optional I/O cable “Cable,CTS, A12-610(standardx2)”
  • Up to 5000 LIFO part results including date, time, barcode, and test results
  • Selectable part serialization in the test data stream
  • SPC data tracking for part traceability
  • Expandable download through USB port
  • 88-264 VAC, 47-63 Hz, or optional 22-26 VDC
  • RS232(2)
  • Telnet(1)
  • Email alerts
  • Barcode ready (RS232)
  • Mobile text
  • USB mass storage device for backup, restore, and reports (FAT32, USB1.0,1.1,2.0)
  • ¼” Colder fitting for test line attachment

The Sentinel I28 leak detection system is configurable for a variety of pressure decay testing or mass flow testing requirements. It provides storage for up to 99 pre-programmed test sequences, and features a “Parent Program” that holds a customized test sequence for your multi-test application.

The test program can be selected via keypad, RS232, or binary. Test result evaluation is user-selectable (fail/pass/fail, etc.). The output format for test results can be configured to meet your needs.

REQUEST A Quote

Leak Test Options

  • Pressure decay/vacuum decay: pressure change
  • Pressure decay/vacuum decay: with certified mass flow leak standard
  • Pressure decay/vacuum decay: pressure change/time
  • Occlusion test (vacuum or pressure test)
  • Mass flow: direct flow reading (pressure test only)
  • Mass flow: continuous direct flow reading (pressure test only)
  • Mass flow: with certified mass flow leak standard (pressure test only)
  • Volumetric fill leak test: IPX6, IP67, sealed part leak test

Selectable Test Parameters

  • Quick test
  • Test pressure level compensation (selectable on/off)
  • Adjustable environmental drift correction per test program